Menu Reference
Calibrate Menu
Click Show Sample of Cal Freqs to display a sample of calibration frequencies from the
current full-sweep calibration.
The CURRENT CAL SETTINGS line shows the Start and Stop frequencies and the number
of points in the current full-sweep calibration.
In the START: and STOP: fields, enter the start and stop frequencies for the new calibration
subset. Changing these fields allow you to sweep a narrower span while the calibration subset
option is active.
In the SKIP PTS: field, enter the number of points to skip in the full calibration to create the
new calibration subset.
The SUBSET CAL SETTINGS line shows the Start and Stop frequencies and the number of
points in the new calibration subset.
Click Create Cal Subset to create the new subset of the current full-sweep calibration set.
134
Chapter 5