Using the Multiport Test System
How to Set Up the Multiport Test System Hardware
Attach the Network Analyzer to the Multiport Test Set
The normal physical location of the test set is beneath the network analyzer in either a
tabletop or rack configuration.
For all test sets except the N4421A, the test set is supplied with notches in the top of the case
to accept the feet of the network analyzer for a secure tabletop installation. Then return to
"How to Set Up the Multiport Test System Hardware" on page 18
Attaching the E8364A/B Network Analyzer to the N4421A Test Set
The E8364A/B network analyzer is attached with the N4421A test set using lock links at the
front and locking feet at the rear. This hardware is supplied with the N4421A test set.
Preparing the Network Analyzer
1. Remove the four feet from the bottom of the network analyzer.
2. Remove the screws from the two lower rear panel standoffs.
3. Install the two rear locking feet where the standoffs were removed. Use part number
Z5823-20239 on the left side of the analyzer and part number Z5823-20240 on the right
side. Use the two longer screws to secure the feet to the analyzer.
Preparing the Test Set
4. Remove the trim strip from the top of the front frame.
20
to complete the installation.