3 - 16
3.3.7

Trigger control via GPIB

You need a separate GPIB command to start a measurement synchronized with
other instruments. This is done by sending the
(Group Execute Trigger) code. The MEASure? and READ? queries do not allow
you to do so, because such a setup causes a query error. With the
INITiate[:IMMediate] and FETCh? concept, it is possible to meet the requirements
of such applications.
Example:
CONFigure:AC
TRIGger:SOURce BUS
INITiate
TRG
*
FETCh:AC?
USING THE COMBISCOPE INSTRUMENTS
Configures the instrument to measure the AC-RMS
voltage.
Specifies that the acquisition is to be triggered by
GET or
TRG.
*
Starts the measurement process.
Triggers the acquisition.
Determines and returns the AC-RMS value.
TRG command or the GET
*