COMMAND REFERENCE
ABORt
Syntax:
ABORt
Description:
The ABORt command resets the trigger system and places it in the "IDLE" state.
Pending actions that were already started are finished immediately. The ABORt
command is not finished until the pending actions have been terminated.
Note:
The commands
functions, except that ABORt does not affect the state of the
INITiate:CONTinuous command. So, when an ABORt command is sent
while the INITiate:CONTinuous is ON, the trigger system will leave the
IDLE state at once.
Example:
Send
ABORt
Send
CONFigure:AC
Send
READ:AC?
<the measured AC-RMS value>
Read
RST and ABORt have the same effect on the trigger
*
Aborts the current acquisition.
Configures for AC-RMS value.
Initiates and reads the AC-RMS value.