COMMAND REFERENCE
SYSTem:SET
Syntax:
SYSTem:SET <indefinite_block>
Query form: SYSTem:SET? [<node_nr> | MINimum | MAXimum]
<node_nr>
Response:
<indefinite_block>
Description:
The SYSTem:SET command programs the instrument to a complete or partial
instrument setup (defined by a node number) using the instrument settings that
were previously retrieved with the SYSTem:SET? query. The instrument settings
are binary settings (bits and bytes) that are changing dynamically. In addition,
various settings are interdependent, even settings divided across different nodes.
For these reasons, instrument settings must not be changed individually.
Appendix E summarizes which instrument settings belong to which node.
If the <node_nr> doesn't exist, the error message -222,"Data out of range;
reserved for future use" is generated. If the <node_nr> is not applicable for this
instrument, the error message -222,"Data out of range; reserved for combi
instrument" is generated.
A number specifying which node settings. The
following nodes are supported:
0
End node indicator.
1|2|3|4 Channel 1 (MINimum) / 2 / 3 / 4 settings
14
Probe scale settings
15
Common vertical settings
16
Horizontal settings
17
Main timebase settings
18
Delayed time base settings
19
Event trigger delay settings
20
SCPI trigger source
32
Cursor settings
33
Cursor autosearch settings
49|50
MEASurement 1/2 settings
51
Pass/Fail test settings
65|66
MATH1/2 settings
80
Display settings
81
Trace intensity settings
82
Display trace position settings
96
Setup label text
112
Autorange settings
128
Real-time clock settings
240
Service (factory) settings (MAXimum)
4 - 105