Fluke CombiScope PM3370B User Manual page 122

Autoranging instrument
Hide thumbs Also See for CombiScope PM3370B:
Table of Contents

Advertisement

4 - 22
RCL
Recall instrument setup
*
Syntax:
RCL <numeric_data>
*
Description:
The
RCL command restores instrument settings from one of the internal memory
*
registers 0 .. 10. The settings in memory register 0 are standard settings, which
can only be recalled. The settings in the memory registers 1 through 10 are
programmable by sending the
After power on the current settings, just before power off, are restored. These
current settings are saved in non-volatile memory (battery backed-up).
Example:
Send
*SAV 2
.
.
Send
*RCL 2
Front panel compliance:
The
SAV/
RCL commands are the remote equivalent of the front panel softkey
*
*
operation via the SETUPS/RECALL menu. The standard settings stored in
memory 0 can be changed via the front panel FRONT SETUPS menu.
SAV command.
*
Stores the actual instrument settings into
memory register 2.
Restores the instrument settings from memory
register 2.
COMMAND REFERENCE