COMMAND REFERENCE
STATus:QUEue[:NEXT]?
Syntax:
STATus:QUEue[:NEXT]?
Response:
<error_number>,"<error_description>"
<error_number>
<error_description> A short description of the error. When there
Description:
The STATus:QUEue? query reports the next event from the error/event queue
and removes this event from the queue. The error queue is a "First-In First-Out"
(FIFO) queue. Therefore, the error query returns the oldest error. Once an error
is read, it is removed from the queue and the next error message is made
available. STATus:QUEue? is the alias of the SYSTem:ERRor? query. If the
queue is empty, the instrument responds with:
0,"No error"
The error/event queue has space for 20 messages. If there are more messages
than the queue can hold, it will overflow. The oldest messages stay in the queue,
but the most recent message is discarded and the latest message is written in its
place. When the event/error queue overflows, the last position in the queue is set
to:
-350,"Queue overflow"
The error/event queue is cleared:
After power on.
When
CLS is received.
*
When the last error in the queue is read.
Example:
Send
STATus:QUEue?
Read
-222,"Data out of range"
The error number is -222 and the meaning is "Data out of range".
A predefined number. If 0 (zero) is returned,
there are no errors in the queue.
are no errors in the queue, the description is
"No error".