Fluke CombiScope PM3370B User Manual page 82

Autoranging instrument
Hide thumbs Also See for CombiScope PM3370B:
Table of Contents

Advertisement

USING THE COMBISCOPE INSTRUMENTS
Example:
Send →
RST
*
Send → DISPlay:MENU MEASure
Send → SYSTem:KEY 2;KEY 4
Send → DISPlay:WINDow:TEXT1:DATA?
Read ← pkpk,6000E-04,V
CURSORS DATA:
The CURSORS function offers a wide variety of voltage and time readouts. The
following readout selections can be made via the CURSORS - READOUT softkey
menu:
<n>:
TYPE:
UNIT:
10
dV
V
dY
U
11
V1
V
12
V2
V
13
Vdc
V
20
dT
s
21
F
Hz
30
dX
U
40
phase
*
51
T1-trg
s
52
T2-trg
s
MATH - FFT DATA:
The MATH1/MATH2 - FFT functions offer the readout of the relative or absolute
frequency and amplitude. The following readout selections can be made via the
CURSORS - READOUT and MATH - FFT - PARAM softkey menus:
<n>:
TYPE:
60
FFT-freq
61
FFT-ampl
DESCRIPTION:
Voltage difference (delta-V) between the cursors.
Vertical voltage (X-deflection on).
Absolute voltage of cursor 1 to ground.
Absolute voltage of cursor 2 to ground.
DC voltage
Time difference (delta-T) between the cursors.
Frequency (1/dT) in Hertz.
Horizontal voltage (X-deflection on).
The phase between two channels in degrees Celsius.
stands for degrees ° sign)
(
*
The time between cursor 1 and the trigger event.
The time between cursor 2 and the trigger event.
UNIT:
DESCRIPTION:
Hz
FFT frequency in Hertz.
variable
FFT amplitude in:
- Relative FFT selected: dB
- Absolute FFT selected: dBm, dbµV, V (Vrms)
'Switches MEAS1 & 2 off
'Switches MEASURE menu on
'Switches MEAS1 and MEAS2 on
'Requests MEAS1 data
'Response = peak-to-peak 0.6 volt.
3 - 63