Fluke CombiScope PM3370B User Manual page 160

Autoranging instrument
Hide thumbs Also See for CombiScope PM3370B:
Table of Contents

Advertisement

4 - 60
INITiate:CONTinuous
Syntax:
INITiate:CONTinuous <Boolean>
Query form: INITiate:CONTinuous?
Response:
1 | 0
1
0
Description:
The INITiate:CONTinuous command selects whether the trigger system is
continuously initiated or not. When INITiate:CONTinuous is ON, the trigger
system is continuously initiating acquisitions. This can only be stopped by setting
INITiate:CONTinuous to OFF or by sending
the current acquisition, but doesn't affect the setting of INITiate:CONTinuous.
Therefore, new acquisitions are initiated immediately.
After a
RST command, INITiate:CONTinuous OFF is valid.
*
PROGRAMMING NOTES:
During INITiate:CONTinuous ON the trigger system remains initiated (does
not return to the IDLE state). This implies that when an
given, bit 0 (OPC) in the standard Event Status Register (ESR) will never be
set. Neither the response to an
a hang-up of the CombiScope when the response is read.
After receiving one of the commands INITiate or INITiate:CONTinuous ON,
the oscilloscope checks the following
X-deflection
Del'd TB
Trigger
Level-PP
X vs Y
Roll mode
Event delay
Peak detection
In case of a settings conflict, the command is ignored and error -221 "Settings
conflict" is reported. To avoid this , first send a
an INITiate command.
Example:
Send
RST
*
Send
CONFigure:AC (@1)
Send
INITiate:CONTinuous ON
Send
FETCh:AC?
Read
<AC-RMS voltage>
Continuous automatic initiation is ON.
Continuous automatic initiation is OFF.
OPC? query will be generated, which causes
*
OFF
OFF
Edge
OFF
OFF
OFF
OFF
OFF
COMMAND REFERENCE
RST. The ABORt command stops
*
RST trigger settings:
*
digital mode
RST command before sending
*
Resets the instrument.
Configures for AC channel 1.
Continuous initiation.
Fetches AC-RMS value.
Reads AC-RMS voltage.
OPC command is
*