Fluke CombiScope PM3370B User Manual page 156

Autoranging instrument
Hide thumbs Also See for CombiScope PM3370B:
Table of Contents

Advertisement

4 - 56
Example 1:
MEASure:VOLTage:AC? 0.6,(@2)
Send
Read
<the measured AC-RMS value>
Send
FETCh:DC? (@2)
Read
<the measured DC component>
Send
FETCh:AMPLitude? (@2)
Read
<the measured amplitude>
Example 2:
Send
CONFigure:AC
Send
TRIGger:SOURce BUS
Send
SENSe:VOLTage:RANGe:OFFSet .25
Send
INITiate
Send
TRG
*
Send
FETCh:AC?
Read
<the measured AC-RMS voltage>
Note:
Because the trigger source is BUS (GPIB), the GET (Group Execute
Trigger) code must be sent after INITiate and before FETCh? to trigger
the acquisition.
Errors:
(1)
When a FETCh? query is executed and no valid acquisition data is
available, nor an acquisition pending, execution error -230, "Data corrupt or
stale" is generated. In that case, no result is returned as response to the
FETCh? query.
(2)
When a FETCh? query for a characteristic from a TRACe memory element
is received, which does not contain valid acquisition data, execution error
-230, "Data corrupt or stale" is generated.
COMMAND REFERENCE
Measures AC- RMS on
channel 2, expected
voltage 600 mV.
Fetches the DC
component.
Fetches the waveform
amplitude.
Configures for AC-RMS.
Trigger source GPIB.
250 millivolt offset.
Initiates trigger system.
Triggers (GET) via GPIB.
Fetches AC-RMS value.