Fluke CombiScope PM3370B User Manual page 220

Autoranging instrument
Hide thumbs Also See for CombiScope PM3370B:
Table of Contents

Advertisement

4 - 120
TRIGger[:SEQuence[1]]:LEVel
TRIGger[:SEQuence[1]]:LEVel:AUTO
TRIGger[:STARt]:LEVel
TRIGger[:STARt]:LEVel:AUTO
Syntax:
TRIGger[:SEQuence[1]]:LEVel <NRf> | MINimum | MAXimum
TRIGger[:SEQuence[1]]:LEVel:AUTO <Boolean>
Alias:
TRIGger[:STARt]:LEVel <NRf> | MINimum | MAXimum
TRIGger[:STARt]:LEVel:AUTO <Boolean>
<NRf>
MINimum
MAXimum Selects the maximum possible trigger level.
Query form: TRIGger[:SEQuence[1]]:LEVel? [MINimum | MAXimum]
Alias:
TRIGger[:STARt]:LEVel? MINimum | MAXimum
Response:
<NR3>
<NR3>
Query form: TRIGger[:SEQuence[1]]:LEVel:AUTO?
Alias:
TRIGger[:STARt]:LEVel:AUTO?
Response:
0 | 1
0
1
Description:
The TRIGger:LEVel command controls the trigger level. The trigger level for the
trigger source is effective only if the trigger source is INTernal 1, 2, 3 or 4. The
instrument function "level-pp" is automatically switched off. If the trigger source is
LINE, execution error -221, "Settings conflict" is generated at receipt of the
command. Execution error -221 is also generated if the instrument cannot report
the unit in volts upon receipt of the query.
The TRIGger:LEVel:AUTO switches the level peak-peak function on or off. If level
peak-peak is switched off, the trigger level is automatically reactivated. If level
peak-peak is switched on, the trigger level is automatically deactivated and the
level range is clamped within the peaks of the signal.
The trigger level expressed in volts.
Selects the minimum possible trigger level.
The trigger level in volts.
Level peak-peak off.
Level peak-peak on.