Fluke CombiScope PM3370B User Manual page 177

Autoranging instrument
Hide thumbs Also See for CombiScope PM3370B:
Table of Contents

Advertisement

COMMAND REFERENCE
SENSe:AVERage:COUNt
SENSe:AVERage:TYPE?
Syntax:
SENSe:AVERage:COUNt <NRf>
<NRf>
SENSe:AVERage:TYPE?
Response:
SCAL
Query form: SENSe:AVERage:COUNt? [MINinum | MAXimum]
Response:
2 | 4 | 8 | 16 | ... | 2048 | 4096
If MINimum was specified, 2 is returned.
If MAXimum was specified, 4096 is returned.
Description:
The SENS:AVER:COUN command sets the preprocessing average count factor.
The count factor is a multiple of 2. The average value is calculated by the
oscilloscope as follows:
AVG
The SENS:AVER:TYPE? query returns the preprocessing average type used,
which is the SCALar implementation.
If the oscilloscope is in the analog mode, error -221 "Settings conflict; Digital
mode required" is generated.
After a
RST command, the average count factor is 8.
*
Example:
Send
*RST
Send
CONFigure:AC
Send
SENSe:AVERage:COUNt 16
Send
SENSe:AVERage ON
Send
INITiate
Send
*WAI
Send
TRACe? CH1
Read
<channel 1 trace averaged over 16 sequential
Front panel compliance:
The SENSe:AVERage:COUNt command is the remote equivalent of the front
panel AVERAGE count option of the ACQUIRE menu.
2 | 4 | 8 | 16 | ... | 2048 | 4096
(
=
X
+
...
n
1
) n ⁄
+
X
n
Resets the instrument.
Configures for AC-RMS.
Average count factor becomes 16.
Switches average function on.
Initiates trace averaging.
Waits for INITiate to finish.
Queries channel 1 trace.
4 - 77
acquisitions>