Fluke CombiScope PM3370B User Manual page 145

Autoranging instrument
Hide thumbs Also See for CombiScope PM3370B:
Table of Contents

Advertisement

COMMAND REFERENCE
DISPlay:BRIGhtness
Syntax:
DISPlay:BRIGhtness <Numeric_data> | MINimum | MAXimum
<Numeric_data>
MINimum
MAXimum
Query form: DISPlay:BRIGhtness? [MINimum | MAXimum]
Response:
<NR3>
<NR3>
Description:
The command sets and the query returns the brightness of the trace display. The
number 0.0 (MINimum) gives the lowest brightness. The number 1.0 (MAXimum)
gives the highest brightness.
Notice that the intensity of text display is not controlled with this command.
After a
RST command, the brightness is set at 1.80E-01, i.e., 0.18.
*
Example:
DISPlay:BRIGhtness 0.5 Sets trace brightness at 0.5.
Send
Front panel compliance:
The DISPlay:BRIGhtness command is the remote equivalent of the front panel
TRACE INTENSITY knob.
0.0 .. 1.0
Equals 0.0 Trace display is fully blanked.
Equals 1.0 Trace display has full intensity.
0.00E00 ... 1.00E00
4 - 45