Fluke CombiScope PM3370B User Manual page 165

Autoranging instrument
Hide thumbs Also See for CombiScope PM3370B:
Table of Contents

Advertisement

COMMAND REFERENCE
INPut<n>:POLarity
Syntax:
INPut<n>:POLarity NORMal | INVerted
<n> 2 | 4
Note:
Query form: INPut<n>:POLarity?
<n> 2 | 4
Response:
NORM | INV
Description:
The INPut<n>:POLarity command sets the polarity of the signal on the input
channels two and four. The polarity can be set to NORMal (default) or INVerted
(inverted signal).
After a
RST command, the polarity is NORMal.
*
Example:
Send
RST
*
Send
CONFigure:AC (@2)
SENSe:FUNCtion "XTIMe:VOLTage2" Sets channel 2 ON.
SendÆ
Send
INPut2:COUPling DC
Send
INPut2:POLarity INVerted
Send
READ:DC? (@2)
Read
<DC value>
Front panel compliance:
The INPut<n>:POLarity command is the remote equivalent of the front panel INV
keys.
Input 4 is not applicable for PM33x0B.
Resets the instrument.
Configures channel 2.
Sets DC input coupling on.
Sets INV CH2 on.
Requests DC channel 2.
Reads inverted DC value.
4 - 65

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents