4 - 66
INSTrument:NSELect
INSTrument[:SELect]
Syntax:
INSTrument:NSELect <NRf> | MINimum | MAXimum
INSTrument[:SELect] DIGital | ANALog
<NRf>
1 | MINimum
2 | MAXimum
DIGital
ANALog
Query form: INSTrument:NSELect? [MINimum | MAXimum}
Response:
1 | 2
1
2
Query form: INSTrument[:SELect]?
Response:
DIG | ANAL
DIG
ANAL
Description:
Selects the analog or digital mode of the CombiScope by specifying a predefined
number or name. When one mode is selected, the other mode is deactivated.
After a
RST command, the digital mode (ANALOG key) is selected.
*
Example:
Send
INSTrument:NSELect 2
Send
INSTrument DIGital
Front panel compliance:
These commands are the remote equivalent of the front panel ANALOG key.
1 | 2
The digital mode (ANALOG key) is activated.
The analog mode is activated.
The digital mode (ANALOG key) is activated.
The analog mode is activated.
The digital mode (ANALOG key) is active. This is
also returned when MAXimum is specified.
The analog mode is active. This is also returned
when MINimum is specified.
The digital mode (ANALOG key) is active.
The analog mode is active.
Analog mode is selected.
Digital mode is selected.