Fluke CombiScope PM3370B User Manual page 215

Autoranging instrument
Hide thumbs Also See for CombiScope PM3370B:
Table of Contents

Advertisement

COMMAND REFERENCE
TRIGger[:SEQuence[1]]:FILTer:HPASs:FREQuency
TRIGger[:STARt]:FILTer:HPASs:FREQuency
TRIGger[:SEQuence[1]]:FILTer:HPASs:STATe
TRIGger[:STARt]:FILTer:HPASs:STATe
Syntax:
TRIGger[:SEQuence[1]]:FILTer:HPASs:FREQuency <NRf>
TRIGger[:SEQuence[1]]:FILTer:HPASs:STATe <Boolean>
Alias:
TRIGger[:STARt]:FILTer:HPASs:FREQuency <NRf>
TRIGger[:STARt]:FILTer:HPASs:STATe <Boolean>
<NRf>
<Boolean> 0 | OFF
Query form: TRIGger[:SEQuence[1]]:FILTer:HPASs:FREQuency?
Alias:
TRIGger[:STARt]:FILTer:HPASs:FREQuency?
Response:
3.00E+04 | 1.00E+08 | 2.00E+08
3.00E+04
1.00E+08
2.00E+08
Query form: TRIGger[:SEQuence[1]]:FILTer:HPASs:STATe?
TRIGger[:SEQuence[1]]:FILTer:HPASs:STATe?
Alias:
Response:
0 | 1
0
1
The cutoff frequency expressed in hertz. The only
possible value is 30000, which defines HF-reject (LF-
pass).
1 | ON
Fixed cutoff frequency of 30 KHz (MINimum).
Bandwidth of 100 MHz
(MAXimum for PM338xB).
Bandwidth of 200 MHz
(MAXimum for PM339xB).
Low-pass filter active.
High-pass filter active (HF-reject).
Sets high-pass filter off.
Sets high-pass filter on.
4 - 115
| MINimum | MAXimum
| MINimum | MAXimum
MINimum | MAXimum
MINimum | MAXimum

Hide quick links:

Advertisement

Table of Contents
loading