4 - 48
DISPlay:WINDow[1]:TEXT<n>:DATA?
Syntax:
DISPlay:WINDow[1]:TEXT<n>:DATA?
[1]
<n> 1 | 2 | 10 | 11 | 12 | 13 | 20 | 21 | 30 | 40 | 51 | 52 | 60 | 61
1
2
10
11
12
13
20
21
30
40
51
52
60
61
Response:
<ASCII_data>
<ASCII_data>
Description:
The DISPlay:WINDow[1]:TEXT<n>:DATA? query returns the measured data as
displayed on the upper line(s) of the screen of your CombiScope instrument.
Indicates that the measurement result field is window 1.
MEAS1 result is returned.
MEAS2 result is returned.
Delta-V/Delta-Y is returned under the following conditions:
TYPE:
ANALOG MODE:
Delta-V
X-deflection off
Delta-Y
X-deflection on
V1 is returned.
V2 is returned.
DC voltage (VDC) is returned.
Delta-T is returned under the following conditions:
TYPE:
ANALOG MODE:
Delta-T
X-deflection off
Frequency (1 / delta-T) is returned.
Delta-X is returned under the following conditions:
TYPE:
ANALOG MODE:
Delta-X
X-deflection on
The phase between 2 channels is returned.
T1-trg is returned.
T2-trg is returned.
FFT frequency in Hz is returned.
FFT amplitude is returned expressed in:
- dB (relative value)
- dBm, dbµV, or Vrms (absolute value)
A sequence of 7-bit ASCII characters.
Example:
Send
Read
Response is peak-peak value of 608.4 mV
(MEAS1).
COMMAND REFERENCE
→
DISPlay:WINDow:TEXT1:DATA?
←
pkpk,6084E-04,V
DIGITAL MODE:
X versus Y off
X versus Y on
DIGITAL MODE:
X versus Y off
DIGITAL MODE:
X versus Y on