4 - 62
INPut<n>:COUPling
Syntax:
INPut<n>:COUPling AC | DC | GROund
<n>
Query form: INPut<n>:COUPling?
<n>
Response:
AC | DC | GRO
Description:
Selects the vertical input coupling of a specified <n> input channel. If AC is
specified, the DC offset value is excluded. If DC is specified, the DC offset value
is included. If GROund is specified, the AC value is grounded (zeroed).
After a
RST command, the coupling is DC.
*
Restrictions:
For the PM33x0B CombiScope instruments channel 3 is not applicable, and the
input coupling of channel 4 (EXT TRIG) can only be set to AC or DC.
Example:
→
Send
RST
*
→
Send
CONFigure:AC (@2)
→
Send
SENSe:FUNCtion "XTIMe:VOLTage2"
→
Send
READ:AC? (@2)
←
Read
<AC-RMS +DC-offset voltage>
→
Send
INPut2:COUPling AC
→
Send
READ:AC? (@2)
←
Read
<AC-RMS voltage>
→
Send
INPut2:COUPling GROund
→
Send
READ:AC? (@2)
←
Read
<AC-RMS ground level voltage>
Front panel compliance:
The INPut<n>:COUPling command is the remote equivalent of the front panel
AC/DC/GND key.
[1] | 2 | 3 | 4
[1] | 2 | 3 | 4
Resets the instrument (DC coupled).
Configures for channel 2 AC-RMS.
Reads AC-RMS on channel 2.
Couples true AC-RMS.
Reads AC-RMS on channel 2.
Couples to ground.
Reads AC-RMS on channel 2.
COMMAND REFERENCE
Sets channel 2 ON.