4 - 42
Example:
→
Send
RST
*
→
Send
CALibration
→
Send
STATus:OPERation:CONDition?
←
Read
<cond_reg>
WHILE (bit 0 of <cond_reg) = 1)
→
Send
STATus:OPERation:CONDition?
←
Read
<cond_reg>
LOOP_WHILE
→
STATus:QUEStionable:CONDition? Requests for questionable
Send
←
Read
<cond_reg>
IF (bit 8 of <cond_reg) = 0)
THEN Calibration_Okay
ELSE (bit 8 of <cond_reg) = 1) Calibration_Not_okay
END_IF
Front panel compliance:
The CALibration command/query is the remote equivalent of the front panel CAL
key.
COMMAND REFERENCE
Resets the instrument.
Starts auto calibration.
Requests for oper. conditions.
Reads condition register.
Loops while calibration busy.
Requests for oper. conditions.
Reads condition register.
conditions.
Reads condition register.