COMMAND REFERENCE
After a
RST command, the trigger level is MAXimum and auto level peak-peak
*
is switched off.
Notice that there exists a coupling between programming the attenuator (vertical
sensitivity) and the trigger level. If the attenuator is changed, the trigger level is
also adapted to keep the signal display on the screen.
Programming tip:
First program the attenuator (SENSe:VOLTage:RANGe:PTPeak), and then the
trigger level (TRIGger:LEVel).
Example:
Send
*RST
Send
TRIGger:SOURce INTernal1
Send
INITiate:CONTinuous ON
Send
SENSe:VOLTage:RANGe:PTPeak 8
Send
TRIGger:LEVel 0.2
Send
TRIGger:LEVel:AUTO ON
Front panel compliance:
The TRIGger:LEVel command is the remote equivalent of the front panel
TRIGGER LEVEL knob. The TRIGger:LEVel:AUTO command is the remote
equivalent of the front panel TRIGGER MAIN TB - level-pp on/off softkey menu.
Resets the instrument.
Trigger
source
channel 1.
Continuous initiation.
1 V/div. sensitivity
Trigger level becomes 0.2 V.
Level
peak-peak
switched off.
Switches level peak-peak on
and deactivates the trigger
level.
4 - 121
becomes
is
also