4 - 44
Example 1:
→
CONFigure:VOLTage:AC 0.6,(@2) Configures AC-RMS channel 2,
Send
→
Send
INPut2:COUPling AC
→
Send
READ:AC? (@2)
←
Read
<first measured AC-RMS value>
→
Send
READ:AC? (@2)
←
Read
<second measured AC-RMS value>
Example 2:
→
Send
CONFigure:VOLTage:RISE:TIME (0.5),20,80,1E-2,(@2)
'
'Configures the rise time, expected voltage 0.5V,
'LOW ref. = 20%,
'HIGH ref. = 80%, expected time 0.01 seconds, channel 2.
'
→
Send
INPut2:COUPling DC
→
Send
READ:RISE:TIME? (@2)
←
Read
<the measured rise time>
→
Send
FETCh:FALL:TIME? (@2)
←
Read
<the measured fall time>
COMMAND REFERENCE
expected voltage 600 mV.
Channel 2 AC coupled.
Initiates + fetches AC-RMS
value.
Initiates + fetches AC-RMS
value.
Channel 2 becomes DC
coupled.
Initiates + fetches the rise time
of the signal on channel 2.
Fetches the fall time of the
signal on channel 2.