COMMAND REFERENCE
INPut<n>:FILTer[:LPASs][:STATe]
INPut<n>:FILTer[:LPASs]:FREQuency?
Syntax:
INPut<n>:FILTer[:LPASs][:STATe] <Boolean>
<n> [1] | 2 | 3 | 4
INPut<n>:FILTer[:LPASs]:FREQuency? [MINimum | MAXimum]
MINimum
MAXimum
Note:
2.00E+07
Response:
Query form: INPut<n>:FILTer[:LPASs][:STATe]?
Response:
0 | 1
0
1
Description:
The INP<n>:FILT command turns the common low-pass filter ON or OFF for all
input channels, independent of the value of <n>.
The INP<n>:FILT:FREQ? query returns the cutoff frequency of the common low-
pass filter, which is fixed at 20 MHz (not programmable). The filter can be turned
ON or OFF with the INPut[<n>]:FILTer[:LPASs][:STATe] command. The common
low-pass filter is called bandwidth limiter on the front panel (BW LIMIT option in the
VERT MENU menu).
After a
RST command, the filter is turned OFF.
*
Example:
Send
INPut:FILTer ON
Front panel compliance:
The INPut<n>:FILTer[:LPASs][:STATe] command is the remote equivalent of the
front panel BW LIMIT option in the VERT MENU menu.
Fixed at 20 MHz
Fixed at 20 MHz
Channel 3 is not applicable for PM33x0B.
Common low pass filter off.
Common low pass filter on.
Turns the filter ON.