4 - 82
SENSe:SWEep:REALtime[:STATe]
Syntax:
SENSe:SWEep:REALtime[:STATe] <Boolean>
Query form: SENSe:SWEep:REALtime[:STATe]?
Response:
0 | 1
0
1
Description:
Switches the 'real- time' mode of the acquisition on or off. If the 'real-time'
sampling mode is switched on, the MTB range is limited to sequential sampling
from 250 nanoseconds through 200 seconds per division (for MTB ranges, refer
to the SENSe:SWEep:TIME command). In the analog mode error -221 "Settings
conflict;Digital mode required" is generated.
After a
RST command, the 'real-time' mode is switched off.
*
Example:
→
Send
*RST
→
Send
SENSe:SWEep:REALtime ON
→
Send
TRIGger:SOURce INTernal1;LEVel .1;SLOPe EITHer
Sets the following trigger settings:
→
Send
INITiate
→
Send
READ:AC?
←
Read
<AC-RMS voltage>
Front panel compliance:
The SENSe:SWEep:REALtime command is the remote equivalent of the front
panel REALTIME ONLY option of the TB MODE menu.
Real-time mode switched off.
Real-time mode switched on.
COMMAND REFERENCE
Resets the instrument.
Sets real-time sampling on.
source = internal channel 1
level = 0.1V
slope = either pos. or neg.
Initiates a single acquisition.
Reads AC-RMS.