Description:
The TRIGger:VIDeo:LINE command selects the video line number. Depending on
the video system selected, the following ranges are valid:
> NTSC
> PAL or SECAM from 1 to 625
> HDTV
The TRIGger:VIDeo:SSIGnal command selects the video signal polarity.
After a
RST command, video line number 1 and signal polarity POSitive are
*
selected.
Example:
Send
TRIGger:TYPE VIDeo
Send
TRIGger:VIDeo:LINE 123
TRIGger:VIDeo:SSIGnal NEGative Selects
Send
Front panel compliance:
The TRIGger:VIDeo:LINE command is the remote equivalent of the front panel
TRIGGER MAIN TB - LINE NBR softkey menu. The TRIGger:VIDeo:SSIGnal
command is the remote equivalent of the front panel TRIGGER MAIN TB -
pos/neg softkey menu.
from 1 to 525
from 1 to 1250
Selects TV video triggering.
Selects video line number
123.
negative
signal polarity.
4 - 133