COMMAND REFERENCE
SENSe:SWEep:TIME
Syntax:
SENSe:SWEep:TIME <NRf> | MINimum | MAXimum
<NRf>
MINimum
MAXimum
Query form: SENSe:SWEep:TIME? [MINimum | MAXimum]
Response:
<NR3>
<NR3>
If MINimum was specified, the minimum possible value is returned.
If MAXimum was specified, the maximum possible value is
returned.
Description:
This command sets the time base of a sweep for all input channels in seconds.
The time base of a sweep is the time duration of one complete trace acquisition.
Together
with
SENSe:SWEep:TIME command determines the Main Time Base (MTB). The
MTB is expressed in seconds per division. Since there are 50 points in each
division (horizontally), the MTB can be calculated from the following equation:
MTB = 50
In the analog mode the main time base is put in the variable mode. In the digital
mode the sweep times are limited by permitted MTB values according to the
following table:
s
ms
500
200
200
100
100
50
50
20
20
10
10
5
5
2
2
1
1
Note:
If 2 or more channels are switched on in the real time mode, the time
base range is limited to 10 µ s (non-alternating time base).
Table 4.2
MTB values in the digital mode
The sweep time in seconds.
Selects the minimum possible sweep time.
Selects the maximum possible sweep time.
The sweep time expressed in seconds.
the
number
SENSe:SWEep:TIME / (TRACe:POINts - 1)
*
µ
s
ns
500
500
250
200
200
100
100
50
50
20
20
10
10
5
5
2
2
1
of
trace
points
not valid
in the real
time mode
(TRACe:POINts),