4 - 92
STATus:PRESet
Syntax:
STATus:PRESet
Description:
The PRESet command is used to set the status data structure in such a way, that
device-dependent events are reported at a higher level through the mandatory
part of the status reporting mechanism. The PRESet command affects only the
enable registers and the transition filters for the device-dependent status data
structures. PRESet does not clear any of the event registers.
Note:
Bit 15 of the 16-bit registers in the Status system is not used and remains
zero. Bit 15 always returns zero when reading these registers.
The following table defines the effect of STATus:PRESet:
STATUS REGISTER
OPERation
QUEStionable
Example:
Send
STATus:PRESet
FILTER / ENABLE
ENABle
PTRansition
NTRansition
ENABle
PTRansition
NTRansition
Presets the status registers as indicated above.
COMMAND REFERENCE
PRESET VALUE
0000 hex.
7FFF hex.
0000 hex.
0000 hex.
7FFF hex.
0000 hex.