If an SRQ is generated (acquisition finished), the dT cursor value is read and
printed by sending:
Request to stop or to repeat this test (do Repeat.test1 again).
Routine ServReq does the following:
-
Serial polls the status byte to reset the SRQ mechanism.
-
Reads the ESR byte to clear the OPC bit.
-
Sets the SRQ.detected flag to signal that an SRQ interrupt occurred.
Routine Enter.Setup does the following:
-
Requests for a path/directory/file_name.
-
Inputs the instrument settings (<setupout$>) from the file specified.
-
Sends the SYSTem:SET <setupout$> command to restore the instrument
setup.
Routine Save.Setup does the following:
-
Requests for a path/directory/file_name.
-
Sends the SYSTem:SET? query and reads in response the <setupin$>
instrument setup.
-
Writes the instrument settings (<setupin$>) to the file specified.
Application program:
Note:
The program is supplied on floppy under file name EXAPPA32.BAS.
APPLICATION PROGRAM EXAMPLES
DISPlay:WINDow:TEXT20:DATA?