Hardware Information
Setting the 6-bit Configuration Switch

Special Settings

Boot-Resident/Main
Firmware update procedures may require this mode in case of firmware
loading errors (main/resident firmware part).
If you use the following switch settings and power the instrument up again,
the instrument firmware stays in the resident/main mode. In resident mode, it
is not operable as a module. It only uses basic functions of the operating
system for example, for communication. In this mode the main firmware can
be loaded (using update utilities).
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode with
default parameter settings.
• Boot Main System / Revert to Default Data
• Boot Resident System / Revert to Default Data
Loss of data
C AU T I O N
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are calibration settings, diagnosis and repair log books which will not be
erased.
If you use the following switch settings and power the instrument up again, it
will start as described above.
168
The instrument will boot to main mode and changes to the module's default
parameter. May be also required to load resident firmware into the module.
The instrument will boot to resident mode and changes to the module's
default parameter. May be also required to load main firmware into the
module.
Save your methods and data before executing a forced cold start.