Installation

Select an Appropriate Location

Install the reader on a level, stable surface in an area where an operating temperature
between 18°C and 40°C can be maintained.
The reader is sensitive to extreme environmental conditions. Avoid:
• Excessive humidity. Condensation directly on the sensitive electronic circuits can
cause the instrument to fail internal self-checks. The humidity must be in the range
described in the Specifications section.
• Excessive ambient light. Bright light may affect the reader's optics and readings,
reducing its linear range.
• Dust. Readings may be affected by extraneous particles in the microplate wells. A
clean work area is necessary to ensure accurate readings.

Install the Power Supply

Power Rating. The instrument's power supply or power cord must be
-WARNING-
connected to a power receptacle that provides voltage and current within
the specified rating for the system. Use of an incompatible power
receptacle may produce electrical shock and fire hazards.
Electrical Grounding. Never use a plug adapter to connect primary power
-WARNING-
to the external power supply. Use of an adapter disconnects the utility
ground, creating a severe shock hazard. Always connect the power cord
directly to an appropriate receptacle with a functional ground.
Power Supply. Use only the power supply shipped with the instrument,
-CAUTION-
and operate it within the range of line voltages listed on it.
1. Connect the power cord to the external power supply.
2. Locate the power inlet on the rear of the reader.
3. Plug the rounded end of the power supply's cord into the power inlet.
4. Plug the other end of the power cord into an appropriate power receptacle.

Prepare the Host Computer

Follow instructions supplied with Gen5 to install the necessary software.
• Ensure the computer meets the minimum system requirements as described in the
Gen5 Instructions for Use.
• You must have administrator privileges to install Gen5. Log in to Windows as
"Administrator" or consult your IT department for assistance.