No Communication (Standard Serial Interface) - Intermec EasyCoder 501 Service Manual

Hide thumbs Also See for EasyCoder 501:
Table of Contents

Advertisement

26.2 Single Function
Failure, cont.
Intermec EasyCoder 501 – Service Manual Ed. 7
No communication (general), cont.
• Often the best way is to enter the Setup Mode via the full
Test Mode described in Chapter 25.4. This action will tempo-
rarily set the parameters of the "uart1:" interface to their default
values and bypass any startup program, allowing you to check
or change the setup values both from the display and via the
serial communication channel "uart1:". In the Test Mode, you
could also read the current setup by printing Test Label #5.
• Make sure that the communication parameters for the printer,
like baudrate, parity control, stop bit protocol, communication
hardware straps, etc., and those of its host computer actually
match.
• In addition, verify that the communication port selected agrees
with the one stated in the Setup Mode.
• With the aid of an oscilloscope, or some kind of line analyser
("break-out box"), verify that the interface signals to and from
the CPU board are correct. The interface circuits could have
been damaged as a result of voltage transients from the inter-
face. Being socket-mounted, these circuits are easily replace-
able.
Repeated damage is, however, an indication of insuffi cient
cable screening, or the electrical environment being too noisy.

No communication (standard serial interface)

• See also above "No communication (general).
• Check that the jumper on connector P-202 of the CPU Board
is correctly inserted so as to give the desired selection between
RS-232, Current Loop, and RS-422, see Chapter 18.
• Check that jumper for DTR/RTS Select on P-200 is placed in
accordance with your cable and protocol.
• Verify that the supply voltages +5V are present.
• If the external device requires current supply from the printer
(+5V), a jumper must be fi tted on connector P-203. Typically,
this applies to the Intermec External Keyboard.
• If Current Loop is selected, verify that optional optocouplers
are mounted in sockets IC-205 and IC-206.
• If RS-422 is selected, verify that an optional RS-422 circuit is
mounted in socket IC-207 and that the communication straps
P-204 – P-207 are fi tted correctly, see Chapter 18.3.
• Using an oscilloscope, verify that all signals during receiving
and transmitting data have the appearance described in this
manual.
Chapter 26 Troubleshooting
205

Advertisement

Table of Contents
loading

Table of Contents