1.2.4.2 Inspection
0007-4603
iR2270 / iR2870 / iR3570 / iR4570
- Checking the Operation of the Leakage Breaker
Advise the user to check the leakage breaker once or twice a month on a regular basis. Be sure also to ask the user
to keep a record of checks.
Inspection Procedure
1) Turn on the main power switch.
2) Push he test button [1] of the leakage breaker with the tip of a ball-point pen [2].
3) Check to make sure that the breaker switch shifts to the OFF side and the power is cut.