4.3.5

Connecting the drain hose

NOTE
It is possible to confuse the suction pipe with the drain pipe.
This could cause waste water to be sucked into the suction machine.
➢ Make sure that the suction and drain pipes are connected properly.
1. Grease the O-ring with the supplied grease (Unisilkon LCA 3801).
2. Insert the end piece of the drain hose (A) (small hose diameter).
The drain is not required if the system is equipped with a central wet
suction system. If a central wet suction system is installed, cover the
drain pipe.
 The drain hose has a continuous slope to allow the waste water
to flow away completely.
4.4

Connecting media (air and water)

4 Installation
4.4 Connecting media (air and water)
31