Configuration
October 2023
6. To complete the calibration, release the button. The calibration process starts and
7. Mount the sensor (see Mounting).
88
Figure 8-29: LED and button location
A. Green LED (OK)
B. Calibration button (SET)
the double flashing of the LED continues for five seconds.
The calibration is finished when the LED stops double flashing, and the light is
constantly green.
Note
If a fault has been detected during the process or if the converter has been already
calibrated with AMS Machine Studio, the calibration is stopped, and the LED flashes
fast (5 Hz) until the AMS EZ 1000 Converter is powered off and on again. The
existing calibration of the converter remains unchanged.
Operating Manual
MHM-97884
MHM-97884, Rev. 1.8