Configuration
October 2023
3. Start AMS Machine Studio.
4. Select the converter to be configured, and click Configure.
5. Enter the configuration parameters according to the used eddy current sensor and
6. Start the calibration process, and select a calibration method (see Calibration).
7. If a single file is needed, save the configuration (Configuration editor → File →
8. Close AMS Machine Studio, and disconnect from the converter.
8.2

Start of an offline configuration

Procedure
1. Select Workspace in the left part of the main view, then click Item → Converter →
64
Figure 8-2: Connection – computer to USB interface
A. USB socket at the computer with AMS Machine Studio installed.
B. USB interface at the converter front.
The AMS EZ 1000 Converter is automatically detected by AMS Machine Studio.
the measuring task (see
Configuration editor and
The configuration is sent to the converter during the calibration process.
The configuration is automatically saved to the current project.
Save as).
The data of the calibration is saved with the configuration file. The data of the
Automatic calibration and the Offline calibration (Easy calibration – calibration
process without AMS Machine Studio) is saved as Manual calibration data.
Note
Before saving the configuration, reload the configuration including the calibration
data from the AMS EZ 1000 Converter. See Reload.
After these steps, the converter is ready for operation.
EZ 1000 converter.
Operating Manual
MHM-97884
parameters).
MHM-97884, Rev. 1.8