9.4

Online commands

After the AMS EZ 1000 Converter is configured and operating, you can issue converter
specific commands to the converter from the Online View.
Note
Ensure that there is an online connection to the converter before using these commands.
Figure 9-16: Converter specific command buttons
A. Calibration Verification Report button
Calibration Verification Report
Use this function to verify the calibration of an already configured and calibrated
measuring chain consisting of an AMS EZ 1000 Converter with connected sensor.
A verification tool such as a CAL 064 with a micrometer screw gauge and a material sample
that matches the material of the measuring object (machine shaft) is required for the
verification.
1. Ensure that the sensor is connected to the converter, the converter is powered, and
2. Mount the sensor into the verification tool.
3. Select Network → EZ 1000
4. Click Calibration Verification Report to open the dialog for verifying the calibration
5. Follow the instructions in the dialog.
100
there is an online connection to the converter.
of the connected AMS EZ 1000 Converter.
Operating Manual
MHM-97884
MHM-97884, Rev. 1.8