Hazardous location installation – application notes
October 2023
Note
Consider this behavior when planning a static distance measurement. When planning a
dynamic distance measurement, ensure that the distance between sensor and measuring
object is adjusted so that a connected multimeter reads a DC voltage of approximately -10
V. This standard adjustment is always recommended (see
at the
Figure 7-5: Static distance measurement – maximum output voltage depending on
U
0
A. Output voltage
B. Nominal output voltage range: -2 V to -18 V
C. Output voltage range with Zener barrier with U
D. Reduced output voltage range with Zener barrier with U
E. Beginning of the measuring range
F. End of the measuring range when using a Zener barrier with U
G. Nominal end of the measuring range and end of the range with a Zener barrier with U
H. Time
Supply boost
The sensor supply boost in the A6500-UM card must be activated or deactivated
depending on the maximum voltage U
configuration of the A6500-UM card where the AMS EZ 1000 Converter with Zener barrier
is connected to and go to Input → Sensor supply boost. See
U
= 26 V
0
U
= 28 V
0
60
machine).
28 V
Remove the checkmark to deactivate the sensor supply boost.
Click the checkbox to activate the sensor supply boost.
Adjustment of measuring chains
= 28 V
0
= 26 V
0
of the selected Zener barrier. Open the
0
Figure
Operating Manual
MHM-97884
= 26 V
0
=
0
7-6.
MHM-97884, Rev. 1.8