Emerson Machinery Health AMS EZ 1000 Operating Manual page 49

Hide thumbs Also See for Machinery Health AMS EZ 1000:
Table of Contents

Advertisement

Operating Manual
MHM-97884
Table 6-3: General
Electrical data
Output and supply circuit
-24 V / ┴/ ┴ / Out terminals
Sensor circuit
Sensor terminal
USB connector
-24 V / ┴/ ┴ / Out terminals
Table 6-4: Capacities and inductances of sensors including connection cable and
extension cable
Sensor
EZ105x, EZ108x, EZ116x,
EZ116H, EZ132x, and PR 642x
with 4 m cable
For each additional meter cable
(maximum total cable length =
30m + 10%)
EZ1900-003-ADAP-1 Adapter
cable for connection of PR 642x
sensors
MHM-97884, Rev. 1.8
With explosion protection class Intrinsically safe, only for
connection to certified intrinsically safe circuits with the
following maximum values.
Open-circuit voltage
U
Sum of short- circuit
ΣI
currents
ΣP
Total power
Within output- and
C
supply circuit effective
L
internal capacity and
inductance
With explosion protection class Intrinsically safe, only for
connection of sensors which are, together with their connection
cable and where applicable their extension cable, do not exceed
the total maximum capacities and inductances.
Open-circuit voltage
U
Short-circuit current
I
Total power
P
C
L
With explosion protection class Intrinsically safe and Increased
safety, only for connection to circuits with the following values.
U
With explosion protection class Increased safety, only for
connection to circuits with the following maximum values.
U
Capacity (C
≤450 pF
+110 pF
+39 pF
= -28 V
i
= -140 mA
i
= 840 mW
i
= 5 nF
i
= 0 mH (negligible)
i
= 7.14 V
0
= 140 mA
0
= 250 mW
0
= 600 nF
0
= 790 µH
0
= 30 V
m
= -60 V
m
)
Sensor
Hazardous location installation
October 2023
Inductance (L
)
Sensor
≤100 µH
negligible
negligible
49