Send And Reload A Configuration; Send A Configuration; Reload A Configuration; Offline (Easy) Calibration - Emerson Machinery Health AMS EZ 1000 Operating Manual

Hide thumbs Also See for Machinery Health AMS EZ 1000:
Table of Contents

Advertisement

Configuration
October 2023
8.6

Send and Reload a configuration

8.6.1

Send a configuration

CAUTION
Connections to external devices may be interrupted when sending configurations.
The output voltage of the AMS EZ 1000 Converter is set to > -1.5 V during this process.
The configuration is generally sent during the calibration process. Use Send & close to
send minor changes to the configuration or configuration files loaded into AMS Machine
Studio to the converter.
Procedure
1. Ensure that there is an online connection between the AMS EZ 1000 Converter and
2. Click Send & close in the ribbon command bar to send the configuration to the AMS
8.6.2

Reload a configuration

Once an online connection has been established, the configuration of the AMS EZ 1000
Converter is automatically loaded to AMS Machine Studio. Click Reload in the ribbon
command bar if the configuration and the calibration data must be loaded again.
8.7

Offline (Easy) calibration

CAUTION
Any work on the system may impair machine protection.
86
AMS Machine Studio.
AMS Machine Studio will automatically establish an online connection to the AMS
EZ 1000 Converter as soon as there is a physical connection through the USB port.
EZ 1000 Converter. The configuration editor automatically closes after the sending
process.
A successfully sent configuration will be indicated by a message in the upper right
corner of the software window. This message window will automatically disappear.
Otherwise close it by clicking on the cross.
The converter is ready to use when the sensor is connected, the converter output
voltage is within the OK range, and the green OK LED on the converter front shows
a steady light.
Note
After sending a configuration the configuration becomes active immediately – this
is indicated by AMS Machine Studio and a steady green LED on the converter front.
The output of the AMS EZ 1000 Converter is activated 25 seconds after the sending
process.
Operating Manual
MHM-97884
MHM-97884, Rev. 1.8