Technical data
October 2023
The incremental scale factor (ISF) error is the maximum amount the scale factor varies
from 16.0 mV per micrometer when measured at specified increments throughout the
linear range. Measurements are taken at 125 µm increments. The ISF error is associated
with errors in radial vibration readings.
Basis of the data are measurements with:
AMS EZ 1000 Converter and EZ190x at the temperature of the AMS EZ 1000 Converter.
First meter of the EZ105x cable at the temperature of the EZ105x tip.
No ambient temperature change of the AMS EZ 1000 Converter during 1 hour before
the measurement. Power and sensor connected to the AMS EZ 1000 Converter for at
least 10 minutes before measurement and calibration.
Emerson reference target material is 42CrMo4 (AISI 4140), unless otherwise stated.
Note
For measurements that require a higher measuring accuracy, factory calibrated measuring
chains are available.
Accuracy of measuring chains different from the standard measuring chain
Add the measurement inaccuracies in
accuracy of measuring chains different from the standard measuring chain. The
incremental scale factor (ISF) error is typically ±5%, this applies to 90% of the measuring
range.
Table 11-4: Additional inaccuracies – DSL
Parameter
0°C to 45°C (Test range)
Cable length 15m
Other target materials
Extended measuring
range: 1.5 mm
(Extended range factor:
1.5)
Manual calibration
110
Table 11-4
to the values in
Additional
Additional
Measurement
Measurement
Inaccuracy (DSL),
Inaccuracy (DSL),
Test range
Operating range
±3%
±4%
±7%
±4%
-7% to +8%
-6% to +5%
2.25%
2.25%
1.75%
2.0%
Operating Manual
MHM-97884
Table 11-3
to get the
Remarks
of nominal value, 5 m
cable length
of nominal value, other
cable length
of nominal value, 4 m
cable length
at +23°C
at -35°C to +85°C
(operating range EZ 1000)
at -35°C to +180°C
(operating range EZ 108x)
Material selected from
AMS Machine Studio
material table
MHM-97884, Rev. 1.8