Emerson Machinery Health AMS EZ 1000 Operating Manual page 87

Hide thumbs Also See for Machinery Health AMS EZ 1000:
Table of Contents

Advertisement

Operating Manual
MHM-97884
The AMS EZ 1000 Converter can be calibrated on EZ108x sensors with 5 m or 10 m sensor
cable without using AMS Machine Studio, provided that the material of the measuring
object is 42CrMo4.
Note
The offline calibration requires an AMS EZ 1000 Converter with factory default settings
(delivery state). If uncertain about the calibration state, use the online view of AMS
Machine Studio (see Details) to check the current calibration type of the converter – Auto,
Easy, or Manual. Factory default settings are also indicated by a continuously triple
flashing of the LED (see
Reset an already configured AMS EZ 1000 Converter to the factory default settings by
using AMS Machine Studio (see
calibration procedure.
An already offline (easy) calibrated AMS EZ 1000 Converter can be offline calibrated again
without resetting the converter to factory default.
Prerequisites
-24 V DC power supply, if the converter is not supplied by a measuring amplifiers such
as an A6500-UM Universal Measuring Card.
AMS EZ 1000 Converter with factory default settings.
EZ108x sensor with 5 m or 10 m sensor cable.
Suitable tool for pressing the calibration button such as a 2 mm screw driver or a pen.
Material of the measuring object: 42CrMo4
Sensor is not installed in the machine.
Procedure
1. Connect the sensor to the AMS EZ 1000 Converter (see
2. Connect the power supply if not already done (see
3. Switch on the power supply.
4. Take the sensor and hold it straight up into the air – away from ferro magnetic
5. With a suitable tool, press and hold the button on the front of the AMS EZ 1000
MHM-97884, Rev. 1.8
Table
3-1).
connection).
Note
Wait 60 seconds after powering up the AMS EZ 1000 Converter or after a firmware
update before starting the calibration process.
materials.
Converter to start the calibration process. The green LED starts flashing. See
8-29
for location of the button and the LED. Hold the button until the LED changes
the flashing pattern to double flashing.
Reset to factory
default) before starting the offline
Configuration
October 2023
Sensor
connection).
Power supply and output signal
Figure
87