The process steps after the selection of the calibration method are described below.
Procedure
1. Enter the calibration data.
78
Use this method for third party sensors, for a configuration
with activated detuning (Measurement → Detune frequency),
or to recalibrate an already calibrated measuring chain.
Depending on the selected sensor (Input → Sensor), a calibration
method is recommended. The recommended method is marked
with (Recommended).
The calibration method is fixed to Multipoint calibration if
33010x is selected for Input → Sensor or the detuning is
activated (Measurement → Detune frequency).
The calibration method Manual calibration cannot be selected
if an EZ105x, an EZ108x, or EZ116x sensor with 1.5 m or 2.0 m
cable length is elected for Input → Sensor.
The calibration method Manual calibration cannot be selected
if PR 6424, PR 6425, or PR 6426 is selected for Input → Sensor.
See
Calibration – PR 6424, PR 6425, and PR 6426 sensors
the recommended calibration procedure if PR 6424, PR 6425,
or PR 6426 sensors are connected to the AMS EZ 1000
Converter.
3. Click Next to continue with the process.
See
Manual calibration – guided
guided
process, or
Multipoint calibration – guided process
details.