Maintenance, fault finding, and repair
October 2023
PC or Laptop with the latest version of AMS Machine Studio installed.
USB cable with Type-A and micro-USB B plug
Latest firmware file
Suitable tool for pressing the calibration button, such as a 2 mm screw driver or a pen.
Procedure
1. Start AMS Machine Studio.
2. Connect the AMS EZ 1000 Converter through the USB interface to the PC.
3. Switch on the power supply of the AMS EZ 1000 Converter, if not already done.
4. In AMS Machine Studio, go to the Advanced tab, and press Ctrl+Alt+M to enable the
5. Select the AMS EZ 1000 Converter to be updated from the device tree.
6. Click Firmware update to open a file browser to browse to the storage location of
7. Select the firmware file, and click Open.
8. Follow the instructions on the screen.
106
AMS Machine Studio automatically establishes a connection to the AMS EZ 1000
Converter.
maintenance mode.
The button for the firmware update appears.
Figure 10-2: Advanced
The button for the firmware update is activated (colored).
the firmware file.
The update process starts. The update dialog opens.
Note
During the update process the output voltage of the converter is set to > -1.5 V DC.
Figure 10-3: Update dialog
Operating Manual
MHM-97884
MHM-97884, Rev. 1.8