Details - Emerson Machinery Health AMS EZ 1000 Operating Manual

Hide thumbs Also See for Machinery Health AMS EZ 1000:
Table of Contents

Advertisement

Online View
October 2023
Table 9-2: Calibration state flags
Flag
Calibration
state
Converter health
This graphic object displays the converter health.
Figure 9-6: Converter health – expanded object
A fault-free converter is indicated with a checkmark within a green solid circle
A faulty converter is indicated with a yellow warning triangle
Click on Converter health to expand the object, and get more information about the
converter health.
Table 9-3: Converter health flags
Flag
Temperature
9.2

Details

Details provides more detailed status information of the converter. The graphic objects
Converter health, Channel OK, Voltage output, and Calibration state are described in
Overview.
94
Meaning
Calibration is OK.
No calibration.
The Easy calibration or
Automatic calibration is not
OK
Meaning
The temperature, measured
by the internal sensor is
within the OK range of -35°C
to 85°C.
The temperature, measured
by the internal temperature
sensor, is out of the OK
range of -35°C to 85°C
Operating Manual
Action
---
Calibrate the converter.
Repeat the Easy or Automatic
calibration. If this is not successful,
calibrate the converter with the
Manual
Manual calibration – guided
process
or Multipoint
Multipoint
calibration – guided process
calibration.
.
Action
---
Take appropriate measures to reduce
the environmental temperature.
MHM-97884, Rev. 1.8
MHM-97884
.