6 Gas Control (Optional)
5. L-shape tubing adapter (x2)
Use the adapter to connect TUBE AA and AB to the threaded female gas
connectors of the VICTOR Nivo reader.
6.2.3

Gas Controller (Rear View)

1. Power Input, DC 12 V (from power supply)
2. Control Output: For digital output and analog input signals processing (not in
use).
3. Alarm: External contact
4. Pure CO
(TUBE A, see page 39).
5. Pure N
(TUBE A, page 39). Leave disconnected, if only CO
40
Danger of Gas Leakage!
Do not expose the tubing to any mechanical or thermal stress.
l
Do not kink the tubing and avoid a small bending radius.
l
If the tubing has to be laid on the floor, provide a corresponding
l
protection to prevent people from stumbling or stepping on the tubing.
Do not lay the tubing close to radiators or other heated objects.
l
Do not expose the tubing to direct sunlight.
l
Check the tubing for damages each time before using gas control.
l
Do not operate the instrument with a damaged gas supply tubing.
l
Contact PerkinElmer to request a new tubing.
Do not manipulate the gas tubing or fittings.
l
The tubing may only be shortened using an approved hose cutter.
l
Input: Push to fit input connector for 6 mm O.D. rigid tubing
2
Input: Push to fit input connector for 6 mm O.D. rigid tubing
2
is used.
2
VICTOR Nivo – User Manual