5 Dispenser Module (Optional)
4. Click Wash/Prime as often as needed.
Notice
Careful and thorough routine washing will ensure a clean dispenser and ideal
l
working conditions. If in doubt, tend to clean with more liquid and more often,
than cutting back on this maintenance step.
If you are only dispensing water or a light buffer, you do not need to follow the
l
above washing protocol each time you are dispensing. It is enough to wash it 2
or 3 times with water (in case of a buffer), and remove the remaining water
afterwards. Still, if the dispenser is used regularly, a weekly wash following the
protocol should be performed.
Solutions which contain proteins will compromise your microplate dispenser
l
performance over time. In that case we recommend rinsing the system daily
with an enzyme detergent solution followed by pure water. On a weekly or
monthly basis, flush with 0.1-0.5 M sodium hydroxide (NaOH), followed by
neutralization with an equivalent normality (0.1-0.5 M) of hydrochloric acid
(HCl). Rinse with pure water to remove the HCl.
Remove the stainless steel weights on the input tubings as well during the
l
cleaning process.
Notice
Washing can also be performed by using the buttons located at the dispenser unit:
The PUMP SEL button is used to select the pump. Green light indicates the
l
selected pump.
The WASH/PRIME button will start the washing of the pump. Priming does not
l
start if the dispensing tip is inside the reader.
5.3

Empty Syringe

After dispensing, the remaining liquids can be returned to the reagent bottle by
clicking the Empty syringe button.
34
VICTOR Nivo – User Manual