6 Gas Control (Optional)
6.2.5
Installing Tubes Between Gas Controller and VICTOR
Nivo reader
1. Screw an L-shape adapter into each of the reader's two gas connectors.
2. Use TUBE AA to connect the Gas Output of the gas controller to the lower gas
connector of the reader. Push the tube into the L-shape adapter.
3. Use TUBE AB to connect the Gas sensors Input of the gas controller to the
upper gas connector of the reader. Push the tube into the L-shape adapter.
6.3

Setting the Pressure

1. Adjust the pressure gauges directly at your CO
pressure of 2.0 bar.
2. Adjust the supplied pressure gauges (2 in the figure above, installed between
the CO
42
Make sure to push the tubes securely all the way into the connectors
avoiding any gas leak. To remove tubing from push-in connectors,
push the red rubber ring while pulling the tubing. If the tubing does
not easily come out, do not force it and simply make sure the red ring
is properly pushed.
/N
tank and the gas controller) to an operating pressure of 0.5 bar:
2
2
and N
tanks (1) to an output
2
2
VICTOR Nivo – User Manual