Dispenser (Optional); Alpha Laser; Electrical Requirements - PerkinElmer VICTOR Nivo User Manual

Hide thumbs Also See for VICTOR Nivo:
Table of Contents

Advertisement

Property
FI detection limit (FITC)
FI dynamic range (FITC)
FP standard deviation (FITC)
TRF detection limit (Europium)
TRF dynamic range
LUM detection limit (ATP)
LUM dynamic range
ALPHA sensitivity
10.1.5

Dispenser (Optional)

Property
Well compatibility
Volume steps (minimum - maximum)
Dispensing
Syringe volume
Dispensing speeds
Accuracy
Precision
Dead volume
10.1.6

Alpha Laser

Property
Laser type
Center wavelength (centroid)
CW power output
Wavelength temperature coefficient
Laser class
10.1.7

Electrical Requirements

One individual power socket is required for each system component:
VICTOR Nivo – User Manual
Value
Top < 0.5 fmol/well (96-well plate)
Top < 0.01 fmol/well (384-well plate)
Bottom < 0.06 fmol/well (384-well plate)
5 orders of magnitude
< 3 mP (96- and 384-well plate)
< 2.3 amol/well (96-well plate)
< 0.5 amol/well (384-well plate)
5 orders of magnitude
< 50 amol (96-well plate)
6 orders of magnitude
< 100 amol (96-well & 384-well plate)
Value
1 – 384 well
1 – 350 µl
2 syringe pumps and injectors
500 µl
5 different (33/67/100/300/500 µl/s)
2% @ 100 µl
1% @ 100 µl
< 650 µl not using back-flush option
Value
Fiber coupled diode laser
680 ± 3 nm @ 350 mW
400 mW
< 0.3 nm/°C
3B
10 Specifications
67

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hh3500Hh35000700Hh35000800

Table of Contents