Barcodes 5, 6 and 7 have two options. You select only one each of these options to install.
NO TE
If, at a later date, you want to change any of these options, you must reinstall all seven
commands in the correct sequence.
1 To read a barcode, point the barcode reader at the barcode and press the

Barcode Requirements

For reliable and accurate reading of barcodes, observe the following
requirements:
• Line thickness must be between 0.19 and 1 mm.
• Maximum width of the barcode is 60 mm. A 60 mm barcode with 1 mm line
• A clear protecting foil is allowed.
• For end-mounted barcodes, the top edge of the bars must be 9 mm or more
Use the Barcode Reader Test function (see
page 79) to check the readability of your barcodes.
1100 Series Sample Capacity Extension
trigger. An audible note confirms that the barcode has been registered.
thickness must be placed on the landscape side of the plate.
above the bottom edge of the plate.