6
Repair Guide
Instructions
Step 4:
Install and tighten 4 screws 0515-1940 diagonally using T8
torque driver with 5lbs-in.
Step 5:
Install Flat Ribbon Cable, 8121-1076 on the board,
N1914-66509 at position P101 .
Step 6:
Install carrier board to mother board at the position shown.
114
Visual
Keysight N1913B/N1914B EPM Series Power Meters Service Guide