U2000 Series Settling Time - Keysight N1913B EPM Series Service Manual

Power meters
Table of Contents

Advertisement

For Keysight U2000 Series power sensors
In FAST mode (using Free Run trigger), for a 10 dB decreasing power step, the
settling time is 25 ms
Table 1-9
Number of
averages
Settling time
(Normal mode)
Settling time
(x2 mode)
[a] Manual filter, 10 dB decreasing power step (not across the switching point)
[1] When a power step crosses the auto-range switch point of the sensor, add 25 ms.
Keysight N1913B/N1914B EPM Series Power Meters Service Guide
[1]
.

U2000 Series settling time

1
2
[a]
(s)
0.045 0.09
0.17
[a]
(s)
0.042 0.05
0.09
Specifications and Characteristics
4
8
16
32
0.34
0.66
1.3
0.17
0.34
0.66
64
128
256
512
2.6
5.2
10.4
20.9
1.3
2.6
5.2
10.4
1
1024
41.9
20.9
39

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1914b epm series

Table of Contents